Description
Details
Features
- Dust and drip proof construction. (designed to IEC 60529 IP54)
- Designed to meet IEC 61010-1 and IEC 61557 safety standard.
- 1mA rated test current at the minimum resistance.
- 200mA measuring current on continuity testing.
- Automatic discharge of circuit capacitance.(Any charge stored in the circuit under test will be automatically discharged after testing.)
- Live circuit warning buzzer and neon lamp.
- Small and lightweight. Shock resistant new case material.
- AC voltmeter with linear, easy-to-read scale.
- Operates on AA, R6P×6 dry batteries.
Specifications:
| Insulation Resistance | |
| Test Voltage | 250V/500V/1000V | 
| Measuring Ranges | 100MΩ/200MΩ/400MΩ | 
| (Mid-scale Value) | (1MΩ) (2MΩ) (4MΩ) | 
| Output Voltage | Rated test voltage+20%, -0% | 
| on open circuit | |
| Nominal Current | 1mA DC min. | 
| Output Short | 1~2mA DC | 
| Circuit Current | |
| Accuracy | 0.1~10MΩ/0.2~20MΩ/0.4~40MΩ | 
| (Accuracy Guaranteed Ranges) ±5% of indicated value | |
| Continuity | |
| Measuring Ranges | 3Ω/500Ω | 
| (Mid-scale Value) | (1.5Ω) / (20Ω) | 
| Output Voltage | 4.1V DC approx. | 
| on open circuit | |
| Measuring Current | 210mA DC min. | 
| Accuracy | |±1.5% of scale length | 
| AC Voltage | |
| AC Voltage | 0~600V AC | 
| Accuracy | |±5% of scale length | 
| General | |
| Safety Standard | IEC 61010-1 CAT.III 600V Pollution Degree 2 | 
| IEC 61010-2-031 | |
| IEC 61557-1/2/4 | |
| IP Rating | IEC 60529-IP54 | 
| EMC Standard | IEC 61326-1 | 
| Withstand Voltage | 3700V AC for 1 minute | 
| Power Source | R6P (AA) (1.5V) × 6 | 
| Dimensions | 106(L) × 160(W) × 72(D)mm | 
| Weight | 560g approx. | 
| Accessories | (Test leads) Pouch for test leads | 
| 8923 (F500mA/600V fuse) × 2 | |
| R6P (AA) × 6 | |
| Shoulder Strap | |
| Instruction Manual | |























 
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
  
 



 
 





 
 